Title of article :
Parameterization of X-ray spectra appropriate for microCT scanners
Author/Authors :
Moya، نويسنده , , U.E. and Brandan، نويسنده , , M.E. and Martيnez-Dلvalos، نويسنده , , Marco A. and Ruiz-Trejo، نويسنده , , C. and Rodrيguez-Villafuerte، نويسنده , , M.، نويسنده ,
Pages :
4
From page :
152
To page :
155
Abstract :
Non-filtered X-ray spectra from three tubes (Oxford Instruments Ultrabright microfocus W anode, Apogee Series 5000 with W and Mo anodes) appropriate for microCT scanners have been measured using a CdTe solid-state detector. The normalized and efficiency-corrected spectra have been parameterized using Boone et al. [10] third-order polynomial expression, obtaining good agreement with the data (typical mean ratio between parameterization and measurement equals 1.02, with standard deviation 0.10). Attenuation of the computed spectra by external filters was analytically simulated, obtaining results that agree well with direct measurements. The air kerma angular distribution of the X-ray beams was measured and the magnitude of the heel effect was evaluated. Tungsten collimators provided by the detector manufacturer had to be used to reduce dead time and it was found that their apertures do not necessarily agree with the nominal values.
Keywords :
Parameterized X-ray spectra , X-ray spectra , microCT
Journal title :
Astroparticle Physics
Record number :
2023537
Link To Document :
بازگشت