Study of DOI resolution and imaging resolution of a PET device
Author/Authors :
Saha، نويسنده , , Lipika and Saitoh، نويسنده , , Kazumi and Kobayashi، نويسنده , , Shigeharu، نويسنده ,
Pages :
8
From page :
27
To page :
34
Abstract :
As a recent trend of DOI measurement for the PET, a simple method of utilizing the light attenuation properties of scintillation materials has been paid attention. We have studied the DOI resolutions for less expensive materials as BGO in both the bench test and the simulation by GEANT4.0. By comparison with both the results, we have recognized the importance of removing the multiple Compton absorption events to obtain the better DOI information. The simulation results for the imaging resolution suggested that its deterioration attributes to the parallax error as well as the systematic displacement inherent in the present method of 3D-reconstruction.
Keywords :
GEANT4 , DOI parameter , Parallax error , Systematic shift , Detection efficiency , PET device