    Attenuation correction factors for cylindrical, disc and box geometry

  • Author/Authors

    Agarwal، نويسنده , , Chhavi and Poi، نويسنده , , Sanhita and Mhatre، نويسنده , , Amol and Goswami، نويسنده , , A. and Gathibandhe، نويسنده , , M.، نويسنده ,

  • Pages
    7
  • From page
    439
  • To page
    445
  • Abstract
    In the present study, attenuation correction factors have been experimentally determined for samples having cylindrical, disc and box geometry and compared with the attenuation correction factors calculated by Hybrid Monte Carlo (HMC) method [ C. Agarwal, S. Poi, A. Goswami, M. Gathibandhe, R.A. Agrawal, Nucl. Instr. and. Meth. A 597 (2008) 198] and with the near-field and far-field formulations available in literature. It has been observed that the near-field formulae, although said to be applicable at close sample-detector geometry, does not work at very close sample-detector configuration. The advantage of the HMC method is that it is found to be valid for all sample-detector geometries.
  • Keywords
    Near-field geometry , Far-field geometry , attenuation correction , High resolution gamma ray spectrometry , HPGe detector , Monte Carlo methods