Title of article :
Simple method for absolute activity measurement of 60Co source
Author/Authors :
Bikit، نويسنده , , Istvan and Nemes، نويسنده , , Tomas and Mrda، نويسنده , , Dusan، نويسنده ,
Pages :
4
From page :
333
To page :
336
Abstract :
The activity of 60Co source was measured using the full absorption, sum and random coincidence (pile-up) peaks and the total spectrum area. It is shown that if the true and random coincidences in the single detector are treated correctly, no additional data are needed for absolute source strength measurement. With the source on the detector end-cap (when the angular correlation effects are negligible), this simple method yields absolute activity values deviating from the reference activity by about 1%.
Keywords :
Sum and random coincidence peak method , Gamma ray spectrometry , Absolute source activity measurement , Coincidence counting
Journal title :
Astroparticle Physics
Record number :
2024822
Link To Document :
بازگشت