Title of article :
A backscatter-suppressed beta spectrometer for neutron decay studies
Author/Authors :
Wietfeldt، نويسنده , , F.E. and Trull، نويسنده , , C. and Anderman، نويسنده , , R. and Bateman، نويسنده , , F.B. and Dewey، نويسنده , , M.S. and Komives، نويسنده , , A. and Thompson، نويسنده , , A.K. and Balashov، نويسنده , , S. and Mostovoy، نويسنده , , Yu.، نويسنده ,
Pages :
18
From page :
574
To page :
591
Abstract :
We describe a beta electron spectrometer for use in an upcoming experiment that will measure the beta-antineutrino correlation coefficient (a coefficient) in neutron beta decay. Electron energy is measured by a thick plastic scintillator detector. A conical array of plastic scintillator veto detectors is used to suppress events where the electron is backscattered. A Monte Carlo simulation of this device in the configuration of the a coefficient experiment is presented. The design, construction, and testing of a full-scale prototype device is described. We discuss the performance of this spectrometer with respect to its suitability for the experiment.
Keywords :
beta decay , Electron spectroscopy , Neutron decay , Scintillation detectors
Journal title :
Astroparticle Physics
Record number :
2025642
Link To Document :
بازگشت