Title of article :
Pulse shaping for Ge-spectrometers optimized for ballistic deficit and electronic noise
Author/Authors :
Kalinin، نويسنده , , A.I. and Bednyakov، نويسنده , , V.A.، نويسنده ,
Pages :
5
From page :
718
To page :
722
Abstract :
For large-volume high-purity Ge detectors, working at low counting rates, a new two-level shaping is proposed which is based on a cusp-like form of the bottom part of the shaping and a parabolic-like form of the top shaping part. Due to the side wings of the cusp the good noise characteristics of the shaping are conserved. At the same time, the parabolic part of the pulse shape allows rather satisfactory compensation of the ballistic deficit. Calculation shows that the noise factor can be improved within 10–12% as compared to standard quasi-Gaussian shaping.
Keywords :
Shaping , Ballistic deficit , Noise , HPGe detectors
Journal title :
Astroparticle Physics
Record number :
2025657
Link To Document :
بازگشت