Title of article :
Low-222Rn nitrogen gas generator for ultra-low background counting systems
Author/Authors :
Wَjcik، نويسنده , , M. and Zuzel، نويسنده , , G.، نويسنده ,
Pages :
6
From page :
427
To page :
432
Abstract :
A new stand-alone system for purification of boil-off nitrogen from 222Rn is described. The purification principle is based on the radon adsorption on charcoal at low temperatures (cryo-adsorption). The device can be attached to ordinary dewars, which are usually used for liquid nitrogen storage or transportation. The measured 222Rn concentration in unpurified boil-off nitrogen is in the range of 0.5 mBq/m3, while the unit described can supply ultra-low background detectors (e.g. gamma spectrometers or dark matter detectors) with gaseous nitrogen of 222Rn specific activity lower by a factor ∼40. The lowest concentration achieved was 12±3 μBq/m3. Using a 100-l dewar it is possible to produce effectively at least 40 m3 of highly pure gas with the available rate between 0.3 and 2.5 m3/h. The nitrogen flux can be set by adjusting the power to an electrical heater placed inside the dewar.
Keywords :
Low background detectors , nitrogen purification , RADON
Journal title :
Astroparticle Physics
Record number :
2025726
Link To Document :
بازگشت