• Title of article

    Low-222Rn nitrogen gas generator for ultra-low background counting systems

  • Author/Authors

    Wَjcik، نويسنده , , M. and Zuzel، نويسنده , , G.، نويسنده ,

  • Pages
    6
  • From page
    427
  • To page
    432
  • Abstract
    A new stand-alone system for purification of boil-off nitrogen from 222Rn is described. The purification principle is based on the radon adsorption on charcoal at low temperatures (cryo-adsorption). The device can be attached to ordinary dewars, which are usually used for liquid nitrogen storage or transportation. The measured 222Rn concentration in unpurified boil-off nitrogen is in the range of 0.5 mBq/m3, while the unit described can supply ultra-low background detectors (e.g. gamma spectrometers or dark matter detectors) with gaseous nitrogen of 222Rn specific activity lower by a factor ∼40. The lowest concentration achieved was 12±3 μBq/m3. Using a 100-l dewar it is possible to produce effectively at least 40 m3 of highly pure gas with the available rate between 0.3 and 2.5 m3/h. The nitrogen flux can be set by adjusting the power to an electrical heater placed inside the dewar.
  • Keywords
    Low background detectors , nitrogen purification , RADON
  • Journal title
    Astroparticle Physics
  • Record number

    2025726