Title of article :
Reconstruction of X-ray spectra with the energy sensitive photon counting detector Medipix2
Author/Authors :
Michel، نويسنده , , Thilo and Talla، نويسنده , , Patrick Takoukam and Firsching، نويسنده , , Markus and Durst، نويسنده , , Jürgen and Bِhnel، نويسنده , , Michael and Anton، نويسنده , , Gisela، نويسنده ,
Pages :
5
From page :
510
To page :
514
Abstract :
A hybrid photon counting detector like the Medipix can be used to measure the energy deposited by X-ray photons in a sensor pixel. We calculate the expected energy deposition distribution with the help of the Monte-Carlo simulation tool ROSI and thus determine the energy response function. These response functions are used in two deconvolution methods to determine X-ray spectra from measured energy deposition spectra. We present reconstruction results for monoenergetic and polychromatic X-ray spectra.
Keywords :
x-ray imaging , Pixel detector , Spectrum reconstruction , photon counting detector , Medipix
Journal title :
Astroparticle Physics
Record number :
2025992
Link To Document :