Title of article :
Edge on silicon microstrip detectors for medical imaging
Author/Authors :
Bergamaschi، نويسنده , , A. and Arfelli، نويسنده , , F. and Dreossi، نويسنده , , D. and Longo، نويسنده , , R. and Olivo، نويسنده , , A. K. Pani، نويسنده , , S. and Rigon، نويسنده , , L. and Castelli، نويسنده , , E.، نويسنده ,
Pages :
6
From page :
199
To page :
204
Abstract :
Silicon microstrip detectors in the edge on geometry, i.e. oriented with the strips parallel to the incoming beam, allow a high absorption efficiency for X-rays in the 10–100 keV energy range. In medical imaging, this results in a significant reduction of the dose given to the patient with respect to conventional screen-film systems. This geometrical configuration was first proposed in the early 90s and since then it has been applied to several medical imaging projects. In this short review, an overview of some medical imaging applications using edge on detectors will be presented, focusing the attention on the main characteristics of the detection systems. The studies performed in Trieste in order to optimize such detectors for digital mammography with synchrotron radiation will be described in detail. In particular, novel imaging techniques (phase contrast, refraction and scatter imaging, computed tomography) implemented and tested at the Trieste synchrotron light source will be shortly discussed.
Keywords :
Single photon counting , Silicon microstrip , Synchrotron radiation , digital mammography
Journal title :
Astroparticle Physics
Record number :
2026583
Link To Document :
بازگشت