Author/Authors :
Ciampolini، نويسنده , , L. and Bove، نويسنده , , L.E. and Mondelli، نويسنده , , C. and Alianelli، نويسنده , , L. and Labbe-Lavigne، نويسنده , , S. and Natali، نويسنده , , F. and Bée، نويسنده , , M. and Deriu، نويسنده , , A.، نويسنده ,
Abstract :
We present new McStas simulations of the back-scattering thermal neutron spectrometer IN13 to evaluate the advantages of a new temperature gradient monochromator relative to a conventional one. The simulations show that a flux gain up to a factor 7 can be obtained with just a 10% loss in energy resolution and a 20% increase in beam spot size at the sample. The results also indicate that a moderate applied temperature gradient ( Δ T ≃ 16 K ) is sufficient to obtain this significant flux gain.