Title of article :
CYBPET: a cylindrical PET system for breast imaging
Author/Authors :
Karimian، نويسنده , , A. and Thompson، نويسنده , , C.J. and Sarkar، نويسنده , , S. and Raisali، نويسنده , , G. and Pani، نويسنده , , R. and Davilu، نويسنده , , H. and Sardari، نويسنده , , D.، نويسنده ,
Pages :
9
From page :
427
To page :
435
Abstract :
We propose a Cylindrical Breast PET (CYBPET) system for breast imaging with patients in the prone position. An individual pendulous breast is covered by thin plastic to provide reduced pressure fixation and surrounded by the crystals inside the CYBPET ring. Each breast is imaged separately. The rest of the body is shielded properly to minimize the contribution of scattered photons from the other breast and the rest of the body. To compare the CYBPET with whole-body PET (WB-PET) the simulations of CYBPET and a WB-PET (GE-Advance) for a 10 mm tumor inside the breast with a lesion to background (breast) activity concentration of 6 to 1 were made. The noise effective count rate (NECR) of CYBPET is about twice that of WB-PET at activity concentrations less than 3.1 μCi/cc. The spatial resolution of CYBPET is better by 25% than the WB-PET.
Keywords :
Positron emission mammography , Positron emission tomography , Nuclear medicine instrumentation , breast imaging , breast cancer , PEM
Journal title :
Astroparticle Physics
Record number :
2026743
Link To Document :
بازگشت