Title of article :
The spectral response of superconducting tunnel junction X-ray detectors
Author/Authors :
Friedrich، نويسنده , , Stephan and Fliegauf، نويسنده , , Rolf and Frank، نويسنده , , Matthias and Veldkamp، نويسنده , , Markus and Labov، نويسنده , , Simon and Beckhoff، نويسنده , , Burkhard and Ulm، نويسنده , , Gerhard، نويسنده ,
Abstract :
The Advanced Detector Group at LLNL is developing superconducting Nb–Al–AlOx–Al–Nb tunnel junction detector arrays for high-resolution soft X-ray spectrometry. We have characterized the detector response function in the energy range between 50 and 2000 eV with monochromatic X-rays, in collaboration with the Physikalisch-Technische Bundesanstalt Laboratory at the BESSY II electron storage ring. The detectors have an energy resolution between 7 and 15 eV FWHM in that energy range, and a count rate capability of ∼10,000 counts/s per detector pixel. In addition, the detector response function exhibits several spectral artifacts, and although their magnitude is only on the order of 1% of the main line, they can reduce the sensitivity for X-ray analysis of dilute specimens if they overlap with the correspondingly weak emission lines. We have characterized these artifacts by scanning a collimated synchrotron radiation beam across the detector area. We identify their origin due to X-ray absorption at locations other than the main Nb absorber film, and outline future detector designs to reduce the artifacts and thereby extend the spectrometer sensitivity.
Keywords :
Superconducting devices , Synchrotron detectors , STJ response functions , Superconducting tunnel junctions , X-ray spectrometers , STJ line shapes
Journal title :
Astroparticle Physics