• Title of article

    Performance study of monolithic line radiation detector based on semi-insulating GaAs using X-ray source

  • Author/Authors

    Za?ko، نويسنده , , B. and Dubeck?، نويسنده , , F. and ??epko، نويسنده , , P. and Melov، نويسنده , , V. and Herms، نويسنده , , M. A. Haupt، نويسنده , , L.، نويسنده ,

  • Pages
    5
  • From page
    78
  • To page
    82
  • Abstract
    The 32-strips linear detector based on semi-insulating GaAs is fabricated. The dimensions of each strip are 170×2500×200 μm3. Analog and digital read-out electronic card with 24 channels is connected to the linear detector. The surface mounted device (SMD) components with automatic assembling of electronic devices for read-out front-end electronics are used. Testing of counting ability of each strip with micro-collimated (φ∼250 μm) X-ray source up to 60 kV is carried out. The maximum counting rate is about 1.0×105 cps. The results for gradual decrease of linearity on approaching the maximum counting rate and the dispersion of the intensity registered in individual channels are presented. The crosstalk between neighbouring strips is found negligible.
  • Keywords
    Semi-insulating GaAs , Strip line detector , X-ray irradiation , digital imaging
  • Journal title
    Astroparticle Physics
  • Record number

    2027247