Title of article :
Count rate performance study of the Lausanne ClearPET scanner demonstrator
Author/Authors :
Rey، نويسنده , , M. and Jan، نويسنده , , S. and Vieira، نويسنده , , J.-M. and Mosset، نويسنده , , J.-B. and Krieguer، نويسنده , , M. and Comtat، نويسنده , , Jean C. and Morel، نويسنده , , C.، نويسنده ,
Pages :
4
From page :
207
To page :
210
Abstract :
This paper presents the count rate measurements obtained with the Lausanne partial ring ClearPET scanner demonstrator and compares them against GATE Monte Carlo simulations. For the present detector setup, a maximum single event count rate of 1.1 Mcps is measured or a 250–750 keV energy window. This corresponds to a coincidence count rate of approximately 22 kcps. Good agreements are observed between measured and simulated data. Count rate performance, including Noise Equivalent Count (NEC) curves, are determined and extrapolated for a full ring ClearPET design using GATE Monte Carlo simulations. For a full ring design with three rings of detector modules, NEC is peaking at about 70 kcps for 20 MBq.
Keywords :
Monte Carlo , Gate , DOI , Small animal imaging , ClearPET , PET
Journal title :
Astroparticle Physics
Record number :
2027317
Link To Document :
بازگشت