Title of article :
Analysis of different detector and electronics defects on F18-FDG images
Author/Authors :
Zito، نويسنده , , Felicia and De Bernardi، نويسنده , , Elisabetta and Schiavini، نويسنده , , Mauro and Canzi، نويسنده , , Cristina and Voltini، نويسنده , , Franco and Agosteo، نويسنده , , Stefano and Gerundini، نويسنده , , Paolo، نويسنده ,
Pages :
5
From page :
493
To page :
497
Abstract :
The effect of malfunctioning detector blocks on accuracy of PET reconstructed images was evaluated qualitatively and quantitatively, by using the ECAT EXACT HR+ scanner. A procedure to simulate defective block was carried out after analyzing and comparing blank scans obtained with detection system in good operation with those having some faulty block detectors. Different levels of detector damages were simulated and tested with phantom studies. The obtained results allowed to better identify type and entity of the artifact expected on reconstructed data and then to know cases of malfunctioning blocks that really need to stop clinical activity.
Keywords :
quality control , PET scanner , Faulty detectors