Title of article
Evaluation of pixellated CZT detectors for neutrinoless double beta-decay measurements
Author/Authors
Bloxham، نويسنده , , T.R. and Freer، نويسنده , , M.، نويسنده ,
Pages
7
From page
722
To page
728
Abstract
The COBRA neutrinoless double beta-decay experiment is currently based in the Gran Sasso underground laboratory. It employs CdZnTe (CZT) detectors which provide both source and detector. Key to these measurements is the reduction of backgrounds to a level at which the system becomes sensitive to half-lives which would potentially allow a determination of the neutrino mass. Here, we evaluate how a pixellated detector with a pixel size approaching 200 μ m would provide a method for enhancing both the experimental signal and suppressing backgrounds present inside the detector volume. It is found that it could be possible to reduce backgrounds from natural activities by up to three orders of magnitude and that neutrinoless double beta-decays can be distinguished from single β -particles with an efficiency of 70 % .
Keywords
Neutrinoless double beta-decay , Pixellated CZT detectors , Background reduction , COBRA-experiment
Journal title
Astroparticle Physics
Record number
2027895
Link To Document