Title of article :
Performance characteristics of a small animal PET camera for molecular imaging
Author/Authors :
Hastings، نويسنده , , D.L. and Reader، نويسنده , , A.J. and Julyan، نويسنده , , P.J. and Zweit، نويسنده , , J. and Jeavons، نويسنده , , A.P. and Jones، نويسنده , , T.، نويسنده ,
Pages :
4
From page :
80
To page :
83
Abstract :
The performance of a novel type of animal PET camera, the quad High-Density Avalanche Chamber (HIDAC) was assessed for a non-rotating 16-module system. Spatial resolution was 1.0 mm, and invariant within a standard deviation ⩽5%. Absolute sensitivity was 0.95%, and the scatter-background corrected sensitivity was 0.75%. The count rate capability was linear at typical activities used in animal imaging, with a 20% loss at 11.5 MBq. The camera demonstrates small regions of radiotracer uptake with excellent detail in the mouse.
Keywords :
Position sensitive detectors , PET , MEDICAL IMAGING
Journal title :
Astroparticle Physics
Record number :
2028008
Link To Document :
بازگشت