• Title of article

    Optimization of a scintillator-deposited charge-coupled device with Monte-Carlo simulation

  • Author/Authors

    Tawa، نويسنده , , Noriaki and Mukai، نويسنده , , Kenji and Ikegami، نويسنده , , Kazuhiro and Miyata، نويسنده , , Emi and Anabuki، نويسنده , , Naohisa and Tsunemi، نويسنده , , Hiroshi and Miyaguchi، نويسنده , , Kazuhisa، نويسنده ,

  • Pages
    3
  • From page
    119
  • To page
    121
  • Abstract
    We report on a new photon-counting detector for X-rays in the 0.1–100 keV energy range. The new detectors are X-ray charge-coupled devices (CCD) on which a scintillator is directly coupled with optical cement. We developed the new detector and measured its spectral resolution. Our results show that the detector possesses a spectral resolution of 26% (FWHM) at 59.5 keV. Additionally, in order to optimize its structure, by employing the DETECT2000 Monte-Carlo simulation software we calculated the light transfer in the scintillator and found the appropriate structure of the scintillator crystal for the new detector.
  • Keywords
    X-Ray , Astronomy , Charge-coupled device , scintillator , Simulation
  • Journal title
    Astroparticle Physics
  • Record number

    2028035