Title of article :
Evaluation of the IR–UV blocking filters for ADR with a TES microcalorimeter
Author/Authors :
Hoshino، نويسنده , , A. and Shinozaki، نويسنده , , K. and Ishisaki، نويسنده , , Y. and Ohashi، نويسنده , , T. and Tanaka، نويسنده , , K. L. Mihara and R. P. Schreiner ، نويسنده , , T. and Tawara، نويسنده , , Y.، نويسنده ,
Abstract :
IR to UV blocking filters are indispensable for refrigerators of space missions, hence material and design of the filters determine overall X-ray detection efficiency. We have tested Toray–Lumirror ( C 10 H 8 O 4 ) and Parylene-N ( C 8 H 8 ) films with thin Al coating, installed in a portable ADR system with a TES microcalorimeter. Aperture cone structures are attached surrounding the films to protect them from adhesion of contaminants. The system was connected to the Manson X-ray generator with a vacuum duct, and the X-ray spectrum from the electron beam target was obtained by cooling down to 100 mK. We observed bremsstrahlung continuum and fluorescence lines from the electron target modified with C- and O-edges of the filters and strong C-/O-K α lines. We fitted the depth of each edge and determined the quantities of both elements in line of sight. The fit result is consistent with the expected values for C and Al, while suggesting 1.3 ± 0.3 μ m build-up of ice for O. We may need to contrive a way to avoid contaminant, e.g., introducing heaters.
Keywords :
Transition edge sensor , X-Ray , Adiabatic demagnetization refrigerator , Transmission , ADR , TES
Journal title :
Astroparticle Physics