• Title of article

    A novel gas-filled detector for synchrotron radiation applications

  • Author/Authors

    Kocsis، نويسنده , , Menyhert and Boesecke، نويسنده , , P. and Carbone، نويسنده , , D. and Herve، نويسنده , , C. and Becker، نويسنده , , B. and Diawara، نويسنده , , Y. and Durst، نويسنده , , R. and Khazins، نويسنده , , D. and He، نويسنده , , B. and Medved، نويسنده , , S. N. Sedov، نويسنده , , V. and Thorson، نويسنده , , T. and Wachter، نويسنده , , G.، نويسنده ,

  • Pages
    5
  • From page
    172
  • To page
    176
  • Abstract
    We describe in this paper the performance under synchrotron radiation of a sealed gaseous detector, Vantec-1, operating at high pressure and less prone to discharges. It is consequently operating at high gain (>105) at high local counting rates up to 106 cps/mm2. To achieve these characteristics, a thin resistive layer is used on a glass plate between the mesh and the readout anode. The performance achieved with this detector is suited for applications in X-ray diffraction and synchrotron radiation. This detector has shown high reliability over time under harsh synchrotron radiation environment and robustness in manufacturing environment.
  • Keywords
    X-ray diffraction , Photon counting , Sealed gaseous detector , Synchrotron radiation
  • Journal title
    Astroparticle Physics
  • Record number

    2029086