Title of article
A novel gas-filled detector for synchrotron radiation applications
Author/Authors
Kocsis، نويسنده , , Menyhert and Boesecke، نويسنده , , P. and Carbone، نويسنده , , D. and Herve، نويسنده , , C. and Becker، نويسنده , , B. and Diawara، نويسنده , , Y. and Durst، نويسنده , , R. and Khazins، نويسنده , , D. and He، نويسنده , , B. and Medved، نويسنده , , S. N. Sedov، نويسنده , , V. and Thorson، نويسنده , , T. and Wachter، نويسنده , , G.، نويسنده ,
Pages
5
From page
172
To page
176
Abstract
We describe in this paper the performance under synchrotron radiation of a sealed gaseous detector, Vantec-1, operating at high pressure and less prone to discharges. It is consequently operating at high gain (>105) at high local counting rates up to 106 cps/mm2. To achieve these characteristics, a thin resistive layer is used on a glass plate between the mesh and the readout anode. The performance achieved with this detector is suited for applications in X-ray diffraction and synchrotron radiation. This detector has shown high reliability over time under harsh synchrotron radiation environment and robustness in manufacturing environment.
Keywords
X-ray diffraction , Photon counting , Sealed gaseous detector , Synchrotron radiation
Journal title
Astroparticle Physics
Record number
2029086
Link To Document