Title of article
Energy resolution and absolute detection efficiency for LSO crystals: A comparison between Monte Carlo simulation and experimental data
Author/Authors
Rothfuss، نويسنده , , Harold and Byars، نويسنده , , Larry C. Casey، نويسنده , , Michael E. and Conti، نويسنده , , Maurizio and Eriksson، نويسنده , , Lars and Michel، نويسنده , , Christian، نويسنده ,
Pages
6
From page
1087
To page
1092
Abstract
Non-proportionality between energy deposition and scintillation light production in a scintillator and transport of scintillation photons in the detector have been introduced in a Geant4-based simulation code. Simulation and experimental results were compared for samples of LSO detectors: absolute detector efficiency and energy resolution obtained from simulation are consistent with the experimental data. We also studied the average path length of scintillation photons in the detector and its contribution to the time resolution.
Keywords
Energy resolution , Detector Efficiency , scintillator , lSO , Non-proportionality , Monte Carlo , Simulation
Journal title
Astroparticle Physics
Record number
2030550
Link To Document