    Calibration of closed-end HPGe detectors using bar (Parallelepiped) sources

  • Author/Authors

    Nafee، نويسنده , , Sherif S. and Abbas، نويسنده , , Mahmoud I.، نويسنده ,

  • Pages
    8
  • From page
    80
  • To page
    87
  • Abstract
    The National Institute of Standards and Technology (NIST, Gaithersburg, MD 20878, USA) uses different extended sources, such as bar (parallelepiped) sources of high activity to calibrate the hyper pure germanium (HPGe) cylindrical detectors. The absolute full-energy peak efficiency and the source self-absorption have been investigated in this work by the use of compact analytical expressions for closed-end HPGe detectors and bar sources. The validity of these expressions is tested through extensive comparisons with the experimental values carried out in the NIST. The comparisons indicate that the direct mathematical method used in the present work is useful in the efficiency calibration of HPGe detectors with bar sources.
  • Keywords
    Bar sources , HPGe detector , Full-energy peak efficiency , Direct mathematical method