Title of article :
Charge collection performance of a segmented planar high-purity germanium detector
Author/Authors :
Cooper، نويسنده , Paul W , R.J. and Boston، نويسنده , , A.J. and Boston، نويسنده , , H.C. and Cresswell، نويسنده , , J.R. and Grint، نويسنده , , A.N. and Harkness، نويسنده , , L.J. and Nolan، نويسنده , , P.J. and Oxley، نويسنده , , D.C. and Scraggs، نويسنده , , D.P. and Lazarus، نويسنده , , I. and Simpson، نويسنده , , J. and Dobson، نويسنده , , J.، نويسنده ,
Pages :
9
From page :
401
To page :
409
Abstract :
High-precision scans of a segmented planar high-purity germanium (HPGe) detector have been performed with a range of finely collimated gamma ray beams allowing the response as a function of gamma ray interaction position to be quantified. This has allowed the development of parametric pulse shape analysis (PSA) techniques and algorithms for the correction of imperfections in performance. In this paper we report on the performance of this detector, designed for use in a positron emission tomography (PET) development system.
Keywords :
Segmented planar HPGe detector , Detector characterisation , Pulse shape analysis , Charge collection
Journal title :
Astroparticle Physics
Record number :
2032832
Link To Document :
بازگشت