Title of article :
Measurement of the thermal conductivity of argon and methane: a test of a transient hot-wire apparatus
Author/Authors :
Pلtek، نويسنده , , J and Klomfar، نويسنده , , J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
17
From page :
147
To page :
163
Abstract :
A transient hot-wire apparatus for absolute measurements of the thermal conductivity of gases has been constructed and tested with argon and methane. The apparatus operates at temperatures from about 300 to 473 K and pressures up to 60 MPa. To evaluate the capability of the apparatus, measurements were performed (i) with argon along five isotherms at temperatures between 300 and 425 K with pressures up to 15 MPa, and (ii) with methane along seven isotherms at temperatures between 290 and 360 K with pressures up to 15 MPa. The experimental thermal conductivity data and the results of their low-density analysis are compared with results of other investigators. The total experimental uncertainty of the measurements is estimated to be ±0.001×p in pressure and ±0.005 K in temperature. The relative uncertainty of the measurements at 95% level is estimated to be 1.2% for argon and 0.7% for methane. The reproducibility of the results measured by the standard deviation from the polynomial λ(ρ) relation along isotherms is 0.3% for argon and 0.2% for methane.
Keywords :
thermal conductivity , Transient hot-wire apparatus , Pure , Methane , argon
Journal title :
Fluid Phase Equilibria
Serial Year :
2002
Journal title :
Fluid Phase Equilibria
Record number :
2033187
Link To Document :
بازگشت