Title of article
An enhanced method to calibrate vibrating tube densimeters
Author/Authors
Bouchot، نويسنده , , C. and Richon، نويسنده , , D.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
20
From page
189
To page
208
Abstract
The conversion of vibrating periods to densities through the calibration of the Anton Paar DMA 512 vibrating tube densimeter (VTD) to perform P–ρ–T measurements on fluids is examined. After reviewing various calibration models with their detailed internal characteristics, a new one is proposed that accurately extends the use of the densimeter over a wide range of densities, including vapors under low pressures, compressed liquids and near critical fluids. A good accuracy is obtained by fitting only two mechanically defined parameters on liquid reference density data. Detailed considerations on the mechanical model and on the calibration procedure are exposed and illustrated through sample results.
Keywords
Vibrating tube , densimeter , Calibration , P–?–T
Journal title
Fluid Phase Equilibria
Serial Year
2001
Journal title
Fluid Phase Equilibria
Record number
2033828