Title of article :
Quantitative amplitude and phase contrast imaging in a scanning transmission X-ray microscope
Author/Authors :
Hornberger، نويسنده , , Benjamin and Feser، نويسنده , , Michael and Jacobsen، نويسنده , , Chris، نويسنده ,
Abstract :
Phase contrast in X-ray imaging provides lower radiation dose, and dramatically higher contrast at multi-keV photon energies when compared with absorption contrast. We describe here the use of a segmented detector in a scanning transmission X-ray microscope to collect partially coherent bright field images. We have adapted a Fourier filter reconstruction technique developed by McCallum, Landauer and Rodenburg to retrieve separate, quantitative maps of specimen phase shift and absorption. This is demonstrated in the imaging of a germanium test pattern using 525 eV soft X-rays.
Keywords :
Configured detector , Phase recovery , Phase contrast , Segmented detector , X-ray microscopy
Journal title :
Astroparticle Physics