Author/Authors :
Yakobson، نويسنده , , B.I. and LaRosa، نويسنده , , A. Duel-Hallen، نويسنده , , H.D. and Paesler، نويسنده , , M.A.، نويسنده ,
Abstract :
The coupling of thermal and optical effects in near-field scanning optical microscope (NSOM) fitted with a coated tapered fiber probe is considered. Heating of the probe occurs due to the absorption of light in the metal coating in the tapered region of the probe at relatively large (tens of microns) distances from the probe apex. Conduction of heat that occurs with a time constant of tens of milliseconds results in an elevated probe temperature over a rather large region of the taper. This in turn, affects the optical throughput of the NSOM. Experimental measurement of thermal/optical coupling and a discussion of coupling mechanisms is included.