Title of article :
A second-order focusing electrostatic toroidal electron spectrometer with 2π radian collection
Author/Authors :
Khursheed، نويسنده , , Anjam and Hoang، نويسنده , , Hung Quang، نويسنده ,
Pages :
7
From page :
104
To page :
110
Abstract :
This paper presents a toroidal electron energy spectrometer designed to capture electrons in the full 2π azimuthal angular direction while at the same time having second-order focusing optics. Simulation results based upon direct ray tracing predict that the relative energy resolution of the spectrometer will be 0.146% and 0.0188% at input angular spreads of ±6° and ±3°, respectively, comparable to the theoretically best resolution of the cylindrical mirror analyzer (CMA), and an order of magnitude better than existing toroidal spectrometers. Also predicted for the spectrometer is a parallel energy acquisition mode of operation, where the energy bandwidth is expected to be >±10% (20% total) of the pass energy. The spectrometer is designed to allow for retardation of the pass energy without the need to incorporate auxiliary lenses.
Keywords :
Auger electron spectroscopy , Scanning electron microscope , Electron energy spectrometers