• Title of article

    Modelling high-resolution electron microscopy based on core-loss spectroscopy

  • Author/Authors

    Allen، نويسنده , , L.J. and Findlay، نويسنده , , S.D. and Oxley، نويسنده , , M.P. and Witte، نويسنده , , C. and Zaluzec، نويسنده , , N.J.، نويسنده ,

  • Pages
    11
  • From page
    1001
  • To page
    1011
  • Abstract
    There are a number of factors affecting the formation of images based on core-loss spectroscopy in high-resolution electron microscopy. We demonstrate unambiguously the need to use a full nonlocal description of the effective core-loss interaction for experimental results obtained from high angular resolution electron channelling electron spectroscopy. The implications of this model are investigated for atomic resolution scanning transmission electron microscopy. Simulations are used to demonstrate that core-loss spectroscopy images formed using fine probes proposed for future microscopes can result in images that do not correspond visually with the structure that has led to their formation. In this context, we also examine the effect of varying detector geometries. The importance of the contribution to core-loss spectroscopy images by dechannelled or diffusely scattered electrons is reiterated here.
  • Keywords
    Core-loss spectroscopy , STEM , Nonlocality , High-resolution electron microscopy , HARECES
  • Journal title
    Astroparticle Physics
  • Record number

    2051923