Title of article :
Platinum electrode functionalized with calix[4]arene thin films for impedimetric detection of sodium ions
Author/Authors :
Sakly، نويسنده , , Souiri، نويسنده , , M and Fekih Romdhane، نويسنده , , F and Ben Ouada، نويسنده , , H and Jaffrezic-Renault، نويسنده , , N، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Pages :
7
From page :
47
To page :
53
Abstract :
A layer of calix[4]arene ionophore deposited on platinum electrode has been characterised by impedance spectroscopy and cyclic voltammetry methods. Ion sensitivity of a device for alkaline ion have been tested for Na+ ion in aqueous solution. The results show, comparatively to EIS and ISFETs devices, a decrease in detection limit (10−8 M) but an increase in response time (30 min). Membrane ageing shows a diminution in coverage rate (50%) but a constant sensitivity for Na+ ion detection.
Keywords :
Platinum , electrochemical impedance , Na+ sensor , coverage rate
Journal title :
Materials Science and Engineering C
Serial Year :
2002
Journal title :
Materials Science and Engineering C
Record number :
2097851