Title of article
Novel sucrose three-enzyme conductometric biosensor
Author/Authors
Soldatkin، نويسنده , , O.O. and Peshkova، نويسنده , , V.M. and Dzyadevych، نويسنده , , S.V. and Soldatkin، نويسنده , , A.P. and Jaffrezic-Renault، نويسنده , , N. and Elʹskaya، نويسنده , , A.V.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
6
From page
959
To page
964
Abstract
For the first time a conductometric biosensor for sucrose determination has been developed using a complex three-enzyme (invertase, mutarotase, and glucose oxidase) containing membrane as a sensitive element immobilized on the conductometric interdigitated planar electrodes. The time of measurement of sucrose concentration in the solution was about 1–2 min. The dynamic range of biosensor depends on buffer capacity, being 2 μM–5 mM of sucrose in 5 mM phosphate buffer. The conductometric biosensor developed demonstrates high selectivity, operational stability and reproducibility. The dependence of sensor response on pH and ionic strength of tested solution has been studied in this work. Storage conditions have also been under investigation. The sensor appeared to be eligible towards application in practice.
Keywords
Invertase , sucrose , mutarotase , Conductometric biosensor