• Title of article

    Wear tests in a hip joint simulator of different CoCrMo counterfaces on UHMWPE

  • Author/Authors

    Gonzلlez-Mora، نويسنده , , V.A. and Hoffmann، نويسنده , , M. F. Stroosnijder، نويسنده , , R. and Gil، نويسنده , , F.J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    153
  • To page
    158
  • Abstract
    The objective in this work was to study the effect of different material counterfaces on the Ultra High Molecular Weight Polyethylene (UHMWPE) wear behavior. The materials used as counterfaces were based on CoCrMo: forged with hand polished and mass finished, CoCrMo coating applied on the forged CoCrMo alloy obtained by Physical Vapour Deposition (PVD). A hip joint simulator was designed and built for these studies. The worn surfaces were observed by optical and scanning electron microscopy. The results showed that the hand polished CoCrMo alloy caused the higher UHMWPE wear of the acetabular cups. The CoCrMo coating caused the least UHMWPE wear, while the mass finished CoCrMo alloy caused an intermediate UHMWPE wear. It is shown that the wear rates obtained in this work are closer to clinical studies than to similar hip joints simulator studies.
  • Keywords
    CoCrMo , WEAR , Hip joint simulator , Metals
  • Journal title
    Materials Science and Engineering C
  • Serial Year
    2009
  • Journal title
    Materials Science and Engineering C
  • Record number