Title of article :
Praseodymium analysis in aqueous solution by Pr3+–PVC membrane sensor based on N,N′-bis(4-hydroxysalicylidene)-1-3-phenylenediamine
Author/Authors :
Zamani، نويسنده , , Hassan Ali and Arvinfar، نويسنده , , Ali and Rahimi، نويسنده , , Fatemeh and Imani، نويسنده , , Alihossien and Ganjali، نويسنده , , Mohammad Reza and Meghdadi، نويسنده , , Soraia، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Abstract :
A new Pr3+ poly vinyl chloride PVC membrane sensor based on a membrane containing 3% N,N′-bis(4-hydroxysalicylidene)-1-3-phenylenediamine (HSPDA) as an ionophore, 2% sodium tetraphenyl borate (NaTPB) as an anionic additive, 65% benzyl acetate (BA) as solvent mediator and 30% poly(vinyl chloride) was prepared. This sensor responds to praseodymium ion in a wide linear dynamic range of 1.0 × 10−6 to 1.0 × 10−2 mol L−1 with Nernstian slope of 19.8 ± 0.4 mV per decade and a detection limit of 5.7 × 10−7 mol L− 1 in pH range of 3.1 to 9.8. It has a fast response time of ~5 s in the whole concentration range, and can be used for at least 2 months without any considerable divergences in the potentials. The proposed sensor displays an excellent selectivity for Pr3+ ions with respect to a large number of alkali, alkaline earth, transition and heavy metal ions. The developed sensor was successfully applied as an indicator electrode in Pr3+ ion potentiometric titration with EDTA, and in direct determination of fluoride ion in two mouth wash samples.
Keywords :
Potentiometry , Ion-selective electrode , Sensor , PVC Membrane
Journal title :
Materials Science and Engineering C
Journal title :
Materials Science and Engineering C