Title of article
Heptadentate Schiff-base based PVC membrane sensor for Fe(III) ion determination in water samples
Author/Authors
Zamani، نويسنده , , Hassan Ali and Ganjali، نويسنده , , Mohammad Reza and Faridbod، نويسنده , , Farnoush and Salavati-Niasari، نويسنده , , Masoud، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
5
From page
564
To page
568
Abstract
Tests performed using heptadentate Schiffʹs base ((tris(3-(thiophenal)propyl)amine (TTA)) toward Fe(III) ions indicated that it could be used as an Fe(III) selective ionophore to be used in a plasticized polymeric membrane sensor. The resulting sensors were found to produce a linear response range of five orders of magnitude with a slope of 19.8 ± 0.3 mV decade− 1 with a detection limit is 8.3 × 10− 9 mol L− 1. The sensor could be used in a pH window of 2.4–4.3 and the response time of the sensor was below 10 s, in addition to its very good Fe(III) selectivity over many mono-,di- and trivalent transition and heavy metal ions. The life time of the electrode was found to be at least 10 weeks.
Keywords
PVC Membrane , Potentiometry , Sensor , Ion-selective electrode
Journal title
Materials Science and Engineering C
Serial Year
2012
Journal title
Materials Science and Engineering C
Record number
2101759
Link To Document