• Title of article

    Application of enzyme/zeolite sensor for urea analysis in serum

  • Author/Authors

    Soldatkin، نويسنده , , O.O. and Kucherenko، نويسنده , , I.S. and Marchenko، نويسنده , , S.V. and Ozansoy Kasap، نويسنده , , B. and Akata، نويسنده , , B. and Soldatkin، نويسنده , , A.P. and Dzyadevych، نويسنده , , S.V.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    6
  • From page
    155
  • To page
    160
  • Abstract
    Urea biosensor based on zeolite-adsorbed urease was applied for analysis of blood serum samples. It should be noted, that this biosensor has a number of advantages, such as simple and fast performance, the absence of toxic compounds during biosensor preparation, high reproducibility and repeatability (RSD = 9% and 4%, respectively). The linear range of urea determination by using the biosensor was 0.003–0.75 mM, and the limit of urea detection was 3 μM. The method of standard addition was used for analysis of serum samples with 500-fold dilution. Total time of analysis was 10 min. Good reproducibility of urea determination in real samples was demonstrated (RSD = 10%). Biosensor results were verified by using a common method of urea determination (diacetyl monoxime reaction). It was shown that by using this biosensor distinguishing healthy people from people with renal dysfunction becomes easier.
  • Keywords
    Silicalite , Urease , Biosensor , Enzyme adsorption , Conductometric transducer
  • Journal title
    Materials Science and Engineering C
  • Serial Year
    2014
  • Journal title
    Materials Science and Engineering C
  • Record number

    2104682