Title of article
Fracture of nickel–titanium superelastic alloy in sodium hypochlorite solution
Author/Authors
Yokoyama، نويسنده , , Ken’ichi and Kaneko، نويسنده , , Kazuyuki and Yabuta، نويسنده , , Eiji and Asaoka، نويسنده , , Kenzo and Sakai، نويسنده , , Jun’ichi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
6
From page
43
To page
48
Abstract
Fracture of the Ni–Ti superelastic alloy for endodontic instruments such as files was investigated with a sustained tensile-loading test in sodium hypochlorite (NaOCl) solution of various concentrations. It was found that the time to fracture was reduced when the applied stress exceeded the critical stress for martensite transformation. When the applied stress was higher than the critical stress, the 0.3 mm diameter wires of the Ni–Ti superelastic alloy sometimes fractured within 60 min. From the results of observations of the fracture surface using a scanning electron microscope, it was revealed that the fracture of the Ni–Ti superelastic alloy is significantly influenced by corrosion when the applied stress was higher than the critical stress for martensite transformation. The results of the present study suggest that one of the causes of the fracture of Ni–Ti files during clinical use is corrosion under the applied stress above the critical stress for martensite transformation in NaOCl solution.
Keywords
Ni–Ti , Superelastic alloy , fracture , Corrosion , Sodium hypochlorite