Title of article :
Optimization of thermal treatment parameters to alter austenitic phase transition temperature of NiTi alloy for medical implant
Author/Authors :
Yeung، نويسنده , , K.W.K. and Cheung، نويسنده , , K.M.C. and Lu، نويسنده , , W.W. and Chung، نويسنده , , C.Y.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
6
From page :
213
To page :
218
Abstract :
Super-elasticity (SE) of nickel titanium alloy is usually achieved by thermal or thermo-mechanical treatment. This SE property depends on reverse martensitic transformation; however, previous studies rarely focused on the effect of heat treatment on the transformation characteristics to the austenite phase. In this study, we aim to explore this relationship because there are numerous developed and potential medical applications utilizing the SE properties of shape memory alloy (SMA) recently. Over 70 specimens heat treated using different treatment parameters were carefully prepared and analyzed by differential scanning calorimeter (DSC). We found that the austenite phase transition temperature (As, Ap and Af) can be manipulated by varying the heat treatment parameters such as heat treatment time, temperature and cooling rate. This is important for the production of NiTi SMA implants with desired SE properties at body temperature.
Keywords :
NiTi shape memory alloy , superelasticity , Reverse martensitic transformation temperatures , Heat treatment
Journal title :
MATERIALS SCIENCE & ENGINEERING: A
Serial Year :
2004
Journal title :
MATERIALS SCIENCE & ENGINEERING: A
Record number :
2146768
Link To Document :
بازگشت