On-Line Measurement of Dissolved Methane Concentration During Methane Fermentation in a Loop Bioreactor
Author/Authors :
-، - نويسنده Biotechnology Group, Department of Chemical Engineering, Faculty of Engineering, Tarbiat Modares University, P.O. Box 14115-143 Tehran, I.R. IRAN Yazdian, Fatemeh , -، - نويسنده Biotechnology Group, Department of Chemical Engineering, Faculty of Engineering, Tarbiat Modares University, P.O. Box 14115-143 Tehran, I.R. IRAN Shojaosadati, Seyed Abbas , -، - نويسنده Biotechnology Group, Department of Chemical Engineering, Faculty of Engineering, Tarbiat Modares University, P.O. Box 14115-143 Tehran, I.R. IRAN Nosrati, Mohsen , -، - نويسنده Biotechnology Group, Department of Chemical Engineering, Faculty of Engineering, Tarbiat Modares University, P.O. Box 14115-143 Tehran, I.R. IRAN Pesaran Hajiabbas, Mahdi , -، - نويسنده Biotechnology Group, Department of Chemical Engineering, Faculty of Engineering, Tarbiat Modares University, P.O. Box 14115-143 Tehran, I.R. IRAN Malek Khosravi, Khosro
Issue Information :
سالنامه با شماره پیاپی 52 سال 2009
Pages :
9
From page :
85
To page :
93
Abstract :
-
Abstract :
A dissolved methane sensor based on silicone tube was designed, constructed and optimized.The silicone tube diameter, silicone tube length and helium flow rate (as the carrier gas(were considered as process parameters to be optimized.A continuous stream of helium (50 mL/min) was directed through the tubing, sweeping out the dissolved methane which diffused through the walls of the tubing from the fermentation broth. The probe was made of silicone rubber tubing, 10 cm in length with inner and outer diameters of 0.25 cm and 0.35 cm, respectively (Detakta Company; NO. 02502). A semi-conductor methane gas sensor (Figaro TGS 2611) - which is highly sensitive and selective to methane gas - was used to measure the dissolved methane continuously. Henry’s law along with a special circuit experimental method was applied for calibration. The output concentration was displayed in mg/l (or ppm) of dissolved gas. The accuracy and response time of this system are ± 2 % and 2 minutes, respectively. Moreover, a control system was installed for recycling methane gas during fermentation
Journal title :
Iranian Journal of Chemistry and Chemical Engineering (IJCCE)
Serial Year :
2009
Journal title :
Iranian Journal of Chemistry and Chemical Engineering (IJCCE)
Record number :
2151969
Link To Document :
بازگشت