Title of article
Optimised cantilever biosensor with piezoresistive read-out
Author/Authors
Rasmussen، نويسنده , , P.A. and Thaysen، نويسنده , , J. and Hansen، نويسنده , , O. and Eriksen، نويسنده , , S.C. and Boisen، نويسنده , , A.، نويسنده ,
Issue Information
دوماهنامه با شماره پیاپی سال 2003
Pages
6
From page
371
To page
376
Abstract
We present a cantilever-based biochemical sensor with piezoresistive read-out which has been optimised for measuring surface stress. The resistors and the electrical wiring on the chip are encapsulated in low-pressure chemical vapor deposition (LPCVD) silicon nitride, so that the chip is well suited for operation in liquids. The wiring is titanium silicide which—in contrast to conventional metal wiring—is compatible with the high-temperature LPCVD coating process.
Keywords
silicide , Cantilever , Piezoresistive read-out , Surface stress