Title of article :
Tension and fatigue behavior of 316LVM 1 × 7 multi-strand cables used as implantable electrodes
Author/Authors :
Lewandowski، نويسنده , , John J. and Varadarajan، نويسنده , , Ravikumar and Smith، نويسنده , , Brian and Tuma، نويسنده , , Chris and Shazly، نويسنده , , Mostafa and Vatamanu، نويسنده , , Luciano O. Mantovano، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
8
From page :
447
To page :
454
Abstract :
The mechanical behavior of 316LVM 1 × 7 cables were evaluated in uniaxial tension and in cyclic strain-controlled fatigue with the use of a Flex tester operated to provide fully reversed bending fatigue. The magnitude of cyclic strains imparted to each cable tested was controlled via the use of different diameter mandrels. Smaller diameter mandrels produced higher values of cyclic strain and lower fatigue life. Multiple samples were tested and analyzed via scanning electron microscopy. The fatigue results were analyzed via a Coffin–Manson–Basquin approach and compared to fatigue data obtained from the literature where testing was conducted on similar materials, but under rotating bending fatigue conditions.
Keywords :
Coffin–Manson–Basquin , S-N , Implantable electrodes , fatigue-life , Stainless steel cables