• Title of article

    A simple and accurate method for calibrating the oscillation amplitude of tuning-fork based AFM sensors

  • Author/Authors

    Liu، نويسنده , , Jinquan and Callegari، نويسنده , , Andrea and Stark، نويسنده , , Martin and Chergui، نويسنده , , Majed، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2008
  • Pages
    4
  • From page
    81
  • To page
    84
  • Abstract
    We have developed a simple and accurate method for calibrating the amplitude of vibration of quartz tuning fork sensors commonly used in atomic force- and near field optical-microscopy. Unlike interferometric methods, which require a complex optical setup, the method we present requires only a simple measurement of the electro-mechanical properties of the tuning-fork oscillator and can be performed in a matter of minutes without disturbing the experimental setup. Comparison with interferometric methods shows that an accuracy of better than few percent can be routinely achieved.
  • Keywords
    NSOM , AFM , Interferometry , Tuning Fork , SNOM
  • Journal title
    Ultramicroscopy
  • Serial Year
    2008
  • Journal title
    Ultramicroscopy
  • Record number

    2157461