Title of article :
A rapid method to correct objective lens astigmatism in a TEM
Author/Authors :
Xing، نويسنده , , Q. and Lograsso، نويسنده , , T.A.، نويسنده ,
Issue Information :
دوماهنامه با شماره پیاپی سال 2009
Pages :
4
From page :
287
To page :
290
Abstract :
This work describes a rapid method to correct the two-fold astigmatism of transmission electron microscope (TEM) objective lens employing caustic curve when no objective aperture is inserted. The method makes use of rounding the caustic curve via the objective lens stigmators after the condenser lens astigmatism has been corrected. It has many advantages over other methods, it is fast, straightforward, and does not need holes or an amorphous material.
Keywords :
Transmission electron microscopy , Caustic curve , Astigmatism
Journal title :
Ultramicroscopy
Serial Year :
2009
Journal title :
Ultramicroscopy
Record number :
2157505
Link To Document :
بازگشت