• Title of article

    Characterization of detector modulation-transfer function with noise, edge, and holographic methods

  • Author/Authors

    McLeod، نويسنده , , Robert A. and Malac، نويسنده , , Marek، نويسنده ,

  • Issue Information
    دوماهنامه با شماره پیاپی سال 2013
  • Pages
    11
  • From page
    42
  • To page
    52
  • Abstract
    We developed a new method for characterization of detector performance used in the transmission electron microscope (TEM) based on the measured contrast of holographic fringes. The new method changes spatial frequency of the measured holographic fringes, generated by an electrostatic biprism and Schottky or cold field-emission gun, to sample the modulation-transfer function (MTF) of the detector. The MTF of a Gatan Ultrascan™ 1000 charged-coupled detector (CCD) is evaluated using the new method and the results are compared to the established noise and slanted-edge method results. Requirements for accuracy of the edge and noise MTF methods are discussed. We consider issues surrounding incomplete read-out and how it affects the gain reference normalization of the detector. We evaluate how the MTF affects optimization of experimental parameters in the TEM.
  • Keywords
    Noise power spectrum (NPS) , Incomplete read-out , Detector quantum efficiency (DQE) , Slow-scan CCD detector , scintillator , modulation transfer function (MTF) , Electron detection , Point spread function (PSF) , Noise transfer function (NTF) , Detector noise , CCD camera
  • Journal title
    Ultramicroscopy
  • Serial Year
    2013
  • Journal title
    Ultramicroscopy
  • Record number