Title of article :
Reduction of multiple hits in atom probe tomography
Author/Authors :
Thuvander، نويسنده , , Mattias and Kvist، نويسنده , , Anders and Johnson، نويسنده , , Lars J.S. and Weidow، نويسنده , , Jonathan and Andrén، نويسنده , , Hans-Olof، نويسنده ,
Issue Information :
دوماهنامه با شماره پیاپی سال 2013
Abstract :
The accuracy of compositional measurements using atom probe tomography is often reduced because some ions are not recorded when several ions hit the detector in close proximity to each other and within a very short time span. In some cases, for example in analysis of carbides, the multiple hits result in a preferential loss of certain elements, namely those elements that frequently field evaporate in bursts or as dissociating molecules. In this paper a method of reducing the effect of multiple hits is explored. A fine metal grid was mounted a few millimeters behind the local electrode, effectively functioning as a filter. This resulted in a decrease in the overall detection efficiency, from 37% to about 5%, but also in a decrease in the fraction of multiple hits. In an analysis of tungsten carbide the fraction of ions originating from multiple hits decreased from 46% to 10%. As a result, the measured carbon concentration increased from 48.2 at%to 49.8 at%, very close to the expected 50.0 at%. The characteristics of the multiple hits were compared for analyses with and without the grid filter.
Keywords :
Detection efficiency , Carbides , tungsten carbide , Quantification , Atom probe tomography , Multiple hits
Journal title :
Ultramicroscopy
Journal title :
Ultramicroscopy