• Title of article

    Detection of X-rays using superconducting series-junction detectors

  • Author/Authors

    Kurakado، نويسنده , , Masahiko and Kirk، نويسنده , , Eugenie C. and Shiki، نويسنده , , Shigetomo and Sato، نويسنده , , Hiromi and Mishima، نويسنده , , Kenji and Otani، نويسنده , , Chiko and Taniguchi، نويسنده , , Kazuo، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    431
  • To page
    436
  • Abstract
    X-rays were measured using quantum superconducting series-junction detectors that consist of an Si substrate and four arrays of superconducting tunnel junctions on the substrate. The detectors are not calorimeters but quantum detectors even though they are making use of phonons; therefore, the detectors can be larger and faster than micro-calorimeters. Superconducting tunnel junctions with thinner tunnel barriers and a new arrangement contributed in obtaining higher energy resolutions than before. The full rise-time of signal outputs from charge-sensitive preamplifiers, i.e., charge collection time of the detector, was approximately 5 μs. A peak-to-background ratio of 8000 and an energy resolution of 70 eV were obtained for 5.9 keV X-rays. The energy resolution is about two times better than those of semiconductor detectors. These results suggest that the superconducting series-junction detectors can be practical high resolution detectors for X-rays.
  • Keywords
    Superconducting tunnel junction , Non-thermal detector , Count rate , 2-D position resolution , Energy resolution , Detection efficiency , Quantum detector
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2010
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number