Title of article :
Image improvement with modified scanning waves and noise reduction in a scanning electron microscope
Author/Authors :
Hwan Kim، نويسنده , , Dong and Jae Kim، نويسنده , , Seung and Kyu Oh، نويسنده , , Se، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
9
From page :
112
To page :
120
Abstract :
In order to acquire a high-resolution image in a scanning electron microscope (SEM), a well-modulated current profile on the deflection coil must be generated by a scan waveform, and a properly selected scan waveform can eliminate the distortion of the image. In this work, various adjustments to the scan waveform, including the frequency, slope, and flatness, were analyzed. A new scan waveform using sinusoidally varying voltages was found to reduce distortion, particularly at the outer range image. The SEM image was also improved by reducing noise from the driver circuit and separating electrical grounding of the scanning coils.
Keywords :
Scanning electron microscope , Scan waveform , Image enhancement , noise reduction
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2010
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2170852
Link To Document :
بازگشت