Title of article :
Diamond detectors for the monitoring of carbon-ion therapy beams
Author/Authors :
Rebisz-Pomorska، نويسنده , , M. and Ciobanu، نويسنده , , M. and Kis، نويسنده , , M. and Pomorski، نويسنده , , M. and Voss، نويسنده , , B.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
6
From page :
534
To page :
539
Abstract :
Over the last few years a significant improvement in the production of single-crystal diamonds produced by Chemical Vapour Deposition (sc-CVD) has been made. Since the availability of the so called ‘detector-grade’ diamond material, we have worked on its application within the framework of the heavy-ion beam therapy. In this paper a study of the sc-CVD detector used for the monitoring of carbon-beam properties is presented. We show the on-line monitoring of the beam flux with an efficiency of (100±2)% for rates up to 108 ions/s for the beam energy of 88 MeV/u. For the energy range of 80–430 MeV/u the monitoring of the beam energy is performed with respective relative resolutions from 3 to 11%. Our results show that a compact on-line beam dose and profile monitor with a required relative precision of 1% percent is feasible with todayʹs technology.
Keywords :
Diamond detectors , Heavy-ion beam monitoring , dosimetry , Single particle counting