Title of article :
ISOLTRAP: a tandem Penning trap system for accurate on-line mass determination of short-lived isotopes
Author/Authors :
Bollen، نويسنده , , G and Becker، نويسنده , , S and Kluge، نويسنده , , H.-J and Kِnig، نويسنده , , M and Moore، نويسنده , , R.B and Otto، نويسنده , , H. Raimbault-Hartmann، نويسنده , , H and Savard، نويسنده , , G and Schweikhard، نويسنده , , L and Stolzenberg، نويسنده , , H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
23
From page :
675
To page :
697
Abstract :
The tandem Penning trap mass spectrometer ISOLTRAP has been set up at the on-line mass separator ISOLDE at CERN/Geneva for accurate mass measurements of short-lived nuclei with T12 ≥ 1 s. The mass measurement is performed via the determination of the cyclotron frequency of an ion in a magnetic field. The design of the spectrometer matches the particular requirements for on-line mass measurements on short-lived isotopes. With the ISOLTRAP spectrometer masses of more than 70 radioactive nuclei have so far been determined with resolving powers exceeding one million and an accuracy of typically 10−7.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2171332