Title of article :
A long-lasting experiment with a 13 g SSG detector
Author/Authors :
Furlan، نويسنده , , M. and Janos، نويسنده , , S. and Schmiemann، نويسنده , , K. and van den Brandt، نويسنده , , B. and Konter، نويسنده , , J.A. and Mango، نويسنده , , S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
3
From page :
17
To page :
19
Abstract :
A SSG detector has been tested in a first long-lasting experiment. It consisted of 13 g of tin granules with diameters of 29±0.5 μm embedded in Teflon powder with a volume filling factor of 11%. The detector has been exposed to cosmic rays and radioactive sources and has been kept for seven weeks at a constant temperature of 70 mK followed by shorter periods at 545 mK and 1.04 K. The detector performance as well as results of rate measurements are presented. These tests were made in preparation of the ORPHEUS dark matter experiment.
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
1996
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2171594
Link To Document :
بازگشت