Author/Authors :
Nagai، نويسنده , , Yasuyoshi and Saito، نويسنده , , Haruo and Iwata، نويسنده , , Tetsuya and Nagashima، نويسنده , , Yasuyuki and Hyodo، نويسنده , , Toshio and Uchida، نويسنده , , Hiroshi and Omura، نويسنده , , Tomohide، نويسنده ,
Abstract :
A new two-dimensional angular correlation of annihilation radiation apparatus is described. Position-sensitive photomultiplier tubes coupled with two-dimensional arrays of small BGO scintillator blocks make simple and compact position-sensitive γ-ray detectors. With a sample-detector distance of 5 m, an angular resolution of 1.1 mrad FWHM and a coincidence count rate of ∼2.4 cps per mCi are obtained. Its performance is demonstrated by the result of a test measurement for KI crystal in which non-localized positronium exists at low temperatures.